• Determination of Mercury at Ultra-Low Levels

Environmental Laboratory

Determination of Mercury at Ultra-Low Levels

Jul 13 2011

With detection limits below 1.0 ppt and linearity of 10ppm the PSA Millennium Merlin range of mercury detectors from PS Analytical (UK) leads the way for the determination of mercury at ultra-low levels. Utilising Cold Vapour Atomic Fluorescence Spectrometry the Millennium range conforms to the most important methodologies for mercury, including EPA 1631, EPA 245.7, ISO 17852, and ASTM 6784-02 and offers an optional AAS detector and speciation capabilities.

A recent application utilised the Millennium range for the analysis of fish and shellfish. Consumption of Hg-containing fish frequently can cause damages to human nerves and organs. The accuracy of the method was validated using a NIST certified mercury reference material. Excellent agreement was obtained with the certified value. The precision on food samples prepared in duplicate was typically below 3%, which is excellent given the complex nature of the samples. For this application the instrument detection limit was found to be 0.2 ng/l. The method detection limit was found to be 0.25 ng/g.
 


Digital Edition

AET 28.2 April/May 2024

May 2024

Business News - Teledyne Marine expands with the acquisition of Valeport - Signal partners with gas analysis experts in Korea Air Monitoring - Continuous Fine Particulate Emission Monitor...

View all digital editions

Events

The World Biogas Expo 2024

Jul 10 2024 Birmingham, UK

ICMGP 2024

Jul 21 2024 Cape Town, South Africa

Australasian Waste & Recycling Expo

Jul 24 2024 Sydney, Australia

Chemical Indonesia

Jul 30 2024 Jakarta, Indonesia

China Energy Summit & Exhibition

Jul 31 2024 Beijing, China

View all events