Environmental Laboratory
Voltammetric Trace Determination of Uranium(VI) in Drinking and Mineral Water According to DIN 38406-17
Apr 02 2008
values range from 10 to 30 ìg/L uranium.
The accurate quantification of trace amounts of uranium generally requires complicated and expensive analytical equipment such as instruments based on inductively coupled plasma mass spectrometry (ICP-MS). In contrast, with the Metrohm 797 VA Computrace, dissolved uranium in the form of U(VI) can be determined easily and at reasonable expenditure! The voltammetric determination of uranium with the 797 VA Computrace is just as accurate and efficient as current ICP-MSmethods. The AdCSVmethod (Adsorptive Cathodic Stripping Voltammetry) for the voltammetric determination of the uraniumcontent is based on complexing U(VI) with chloranilic acid (2,5-dichloro-3,6-dihydroxy-1,4-benzoquinone). The uraniumcontent can be determined down to the ng/L range.
DIN Standard 38406-17 is currently being drawn up based on this AdCSV method. This means that in future the voltammetric determination of uranium can be used as a generally recognised method for monitoring drinking and mineral water in daily practice.
The 705 UV Digester made by Metrohm is used for the UV photolysis of water samples with a low to medium organic load. The UV Digester consists of a control unit with built-in timer and a separate wet-chemistry part with UV lamp, cooling device and a holder for 12 sample digestion tubes, each for max. 12 mL sample.
Digital Edition
AET 28.4 Oct/Nov 2024
November 2024
Gas Detection - Go from lagging to leading: why investment in gas detection makes sense Air Monitoring - Swirl and vortex meters will aid green hydrogen production - Beyond the Stack: Emi...
View all digital editions
Events
Nov 26 2024 Paris, France
Nov 27 2024 Istanbul, Turkey
H2O Accadueo International Water Exhibition
Nov 27 2024 Bari, Italy
Biogas Convention & Trade Fair 2024
Nov 27 2024 Hanover, Germany
Dec 02 2024 London, UK