Environmental laboratory
Metrosep C3 – 250 IC Column with Unique Carrier Material for Improved Peak Separation with Monovalent and Bivalent Cations
Jan 01 2000
The relevant column parameter is referred to as the number of theoretical plates per meter (TP/m).
The Metrosep C3 â 250 reaches 42'000 TP/m for sodium, 51'000 for ammonium and 31'000 for barium, which elutes rather late. The peak shapes obtained with this cation column are perfectly symmetrical and â compared to the silicate carrier material â the separation between sodium and ammonium is significantly improved.
The Metrosep C3's excellent selectivity also allows for the separation of transition metals. As the Metrohm ion chromatographs generally determine cations without chemical suppression, the Metrosep C3 â 250 permits determination of the transition metals together with the alkali and alkaline earth metals.
The column is highly suitable for the separation of Li+, Na+, K+, Rb+, Cs+, Mg2+, Ca2+, Sr2+, Ba2+, NH4+, Mn2+, Co2+, Zn2+, Ni2+ in the lower ppb range, with excellent separation between Na+ and NH4+.
Generally low detection limits and the possibility to determine widely differing concentrations prove highly beneficial. The new Metrosep cation column is made of PEEK, its length is 250 mm and
its internal diameter 4 mm.
Digital Edition
AET 28.4 Oct/Nov 2024
November 2024
Gas Detection - Go from lagging to leading: why investment in gas detection makes sense Air Monitoring - Swirl and vortex meters will aid green hydrogen production - Beyond the Stack: Emi...
View all digital editions
Events
Jan 12 2025 Abu Dhabi, UAE
Jan 14 2025 Abu Dhabi, UAE
Jan 20 2025 San Diego, CA, USA
Carrefour des Gestions Locales de L'eau
Jan 22 2025 Rennes, France
Safety, Health & Wellbeing LIVE
Jan 22 2025 Manchester, UK