Environmental Laboratory
Small Infrared Sources: Powerful and energy efficient
Why smaller can be better
Feb 14 2020
Author: Pekka Luoma on behalf of Axetris AG
The Non-Dispersive-Infrared (NDIR) and Photo-Acoustic-Spectroscopy (PAS) sensor application segments are experiencing fast, world-wide growth. The growth is due to the increasing demand for monitoring and control, as well as the ability to switch from classical methods to infrared based applications when gas selectivity, long life-span and high-accuracy are needed. One of the industry leading trends is to continuously minimize size and optimize the sensor units to serve new market segments. Miniaturization increases the requirements for component manufacturers and demands innovative and customized solutions.
A typical NDIR or PAS setup consists of an infrared light emitter, a gas-specific, selective band pass filter and a pyroelectric, or thermopile detector respectively for PAS a piezo microphone. In various applications, multi channel detectors are built in, either as a reference or to measure several gases. Detectors vary in price and performance, e.g. special detectivity, but are similarly responsive across the whole infrared range.
The sensor units are typically in use for years, and thus a significant amount of time and effort is spent on qualifying and conducting lifetime tests. A clear advantage is to choose a light source suitable for multiple sensor applications. IR Source characteristics for universal usage are high emissivity throughout the wavelength range, high optical power, large modulation depth with higher frequencies, long life-span and compact size (Figure 1 & 2). If these requirements are met, only the band pass filter and the optical path need to be fine-tuned to adapt the sensor design to other gases and applications. Thus, months or years of lifetime tests can be saved.
Sometimes Smaller is Better
Minimizing the size of a MEMS IR emitter comes with benefits and drawbacks. A clear drawback is the smaller, active surface: following the Stefan–Boltzmann law, the emitting power is linearly proportional to the area, and thus reducing the chip size while keeping the temperature profile the same, reduces optical power. The benefits on the other hand, include reduced physical size, increased electrical efficiency and high-modulation depth with higher frequencies.
A clear advantage gained by the smaller chip is the relation of the chip size to reflector outer diameter: the light from a point source can be guided forward with much greater efficiency, especially when considering typical gas cuvette diameters range from two to four millimeters and typical high integrated IR multichannel detector with a small Field-of-View.
Lower Power Consumption with Smaller Sensors
The total power consumption from a smaller sensor is becoming crucial when sensors are portable or battery driven. As the IR emitter influences the energy consumption significantly, efficient emitters with a small active area are required. The Axetris EMIRS technology enables to operate it with a short duty cycle, reducing the power consumption below 10 mW. Additionally, a high time resolution can be achieved by driving the EMIRS50 sources up to 100 Hz. The loss of modulation depth caused by the high frequency can be compensated by increasing the input power respectively voltage (Figure 5). Furthermore, an increase in the signal-to-noise ratio can be achieved by operating the IR Source with a duty cycle below 15% at a very high frequency combined with compensated input power.
High Volume, Small Sensors
Many new, high-volume market segments are only accessible with certain end-user pricing. This sets clear boundary conditions for each supplier. At Axetris, multiple steps have been taken to meet specific needs, and as presented in Figure 6, significant cost-reduction activities have been carried out. Lastly, fully tested SMD packaged and chip level tested products are available to meet future needs.
Small, Thus Fast
Scaling down the size of the emitters means reducing the thermal mass. Cooling and heating is faster, allowing the IR Source to operate with higher frequencies. The modulation depth, i.e. the actual signal from the detector, is then higher above a certain frequency, compared to a source with larger thermal mass.
The requirements for frequency are application-dependent: In capnography for example, the requirements are high. In order to draw useful data for a capnogram, it is vital to acquire 30 to 100 measurement points per second. For this reason, an emitter operating reliably at a high sample rate is necessary. With the EMIRS50, Axetris provides an IR Source capable of emitting remarkably fast: modulation depth exceeds 60% at 100 Hz (Figure 4).
A comparison measurement shown in Figure 7, was completed with the larger Axetris EMIRS200 with Reflector 1 source and smaller EMIRS50 with Reflector 6 to study their performances: Frequency scans were conducted with a 20 mm distance from the end of the reflector to the detector. The larger EMIRS200 has clear advantages in frequency range below 17 Hz. However, when frequency is above 17 Hz, the much smaller EMIRS50 shows clear benefits and more intensity.
Small and Efficient
For a measurement system to provide fast response times, the gas exchange must also be rapid and the space to be filled should be as small as possible. To obtain forced flow through sensors, closed reflective cuvette geometry is often used (Figure 9). The measuring range and lowest detection limit (in function of the gas absorption coefficient), together with the used detector and IR emitter define the length of the cuvette. The diameter of the cuvette should be as small as possible, but it depends on the coupling of the IR emitter and detector.
With Axetris’ customized reflectors, all of the light from a small emitter can be guided to a cuvette with a diameter of 3 mm (Figure 8). The angle distribution is optimized with the shape of the reflector. Since the properties of the band pass filter are dependent on the incoming angle of the light, it is beneficial to either measure the actual performance of the band pass filter with spectrometer or simulate the expected performance.
For compact diffusive sensor designs, multi-pass arrangements (Figure 10) are often used to keep the air volume low and to achieve sufficient gas absorption length. For these kinds of setups, well collimated and focused IR emitters are required to guide most of the IR light through the optical system. The smaller, highly efficient EMIRS50 reflectors offer a significant performance increase to such an advanced and high performing sensor design.
When comparing the performance of the smaller EMIRS50 IR Source with the larger “standard” EMIRS200 IR Source, both with optimized reflector designs, the efficiency increase with the EMIRS50 Source is clear: in gas cuvette simulation (2x2 mm detector with 1 mm distance) with < 8° angle of incident, 38% of the total optical power with the EMIRS50 Source fall in the detector, compared to 8.3 % of the total optical power of the larger EMIRS200 Source (Figure 8). The result is an optical efficiency being practically five times higher with the smaller EMIRS50 IR Source.
Conclusion:
Increasing demand for monitoring, process control and miniaturization of sensors requires new, customized solutions, such as IR Sources, from component manufacturers. Smaller is often faster, more cost effective and less power consuming which is crucial for portable sensors. However, the requirements for an IR emitter depend on the application, on the sensor principle, the optical geometry and many more specific characteristics.
The Axetris team collaborates with its customers to find the right product design that best suits their needs regarding performance and costs. Furthermore, our application team supports our valued customers by carrying out optical analysis and optimization of specific parameters.
Digital Edition
AET 28.4 Oct/Nov 2024
November 2024
Gas Detection - Go from lagging to leading: why investment in gas detection makes sense Air Monitoring - Swirl and vortex meters will aid green hydrogen production - Beyond the Stack: Emi...
View all digital editions
Events
Nov 26 2024 Paris, France
Nov 27 2024 Istanbul, Turkey
H2O Accadueo International Water Exhibition
Nov 27 2024 Bari, Italy
Biogas Convention & Trade Fair 2024
Nov 27 2024 Hanover, Germany
Dec 02 2024 London, UK