• Avoid the Sample Capacity Pitfall when Converting Volatile Methods to Fast GC  
    Figure 1. Conventional GC Analysis, 30:1 Split.
  • Figure 2. Fast GC Analysis, 30:1 Split.
  • Figure 3. Fast GC Analysis, 100:1 Split.

Water/Wastewater

Avoid the Sample Capacity Pitfall when Converting Volatile Methods to Fast GC  

Fast GC is the manipulation of several parameters to provide faster analyses while maintaining resolution. It is accomplished using a short column (reduces analysis time) with a narrow I.D. (offsets the loss of efficiency of the short column) while adjusting specific operating parameters, such as linear velocity and oven temperature ramp rates. This short eBulletin is based on a longer Reporter article that includes several more examples and chromotograms.

Converting a conventional GC method to a Fast GC method is not as simple as just changing to a smaller I.D. column. Column dimensions, linear velocity, and oven temperature ramp rates must be optimised together. Changing only one parameter may decrease analysis time (desirable), but will likely cause a loss of resolution (undesirable). It is also critical to account for the reduced sample capacity of the smaller I.D. column.

A method for the analysis of volatiles from waste water samples on a SPB-624 column was selected to illustrate the change from conventional GC to Fast GC. The optimised chromatogram obtained using conventional GC is shown in Figure 1. The analysis time is <18 minutes, peak shapes are good, and the mass spectrometer (MS) is able to mass resolve all analytes.

The column dimensions were changed to Fast GC dimensions, and then linear velocity and oven temperature ramp rates were optimised. The resulting chromatogram is shown in Figure 2 (parameters which were changed are highlighted). While analysis time is reduced to <10 minutes and the MS is still able to mass resolve all analytes, peak shape overall is not good. Why the poor chromatography? The reduced sample capacity of the smaller I.D. column was not accounted for. The capacity of the 0.18 mm I.D. column is less than the 0.25 mm I.D. column, leading to sample overload.

To alleviate this problem, the mass of sample reaching the column must be reduced. Diluting the sample, decreasing the injection volume, or increasing the split ratio are options to achieve this. The first two options are not as compatible with P&T methods as the third. Therefore, the split ratio was increased from 30:1 to 100:1, and then the linear velocity was again optimised. The resulting chromatogram is shown in Figure 3 (parameters which were changed are highlighted). Now, the short analysis time (<10 minutes) and mass resolution of all analytes is accompanied with improved peak shape of all analytes. This represents a vast improvement in analysis time (from over 17 minutes to under 10 minutes) compared to the conventional GC run, with no decrease in quality (resolution and/or peak shape).

For additional information on this or other aspects of GC, visit the GC Learning Centre, a technical repository of column selection criteria, chromatogram search tools, presentations, on-demand webcasts, and brochures/bulletins.


Digital Edition

AET 28.4 Oct/Nov 2024

November 2024

Gas Detection - Go from lagging to leading: why investment in gas detection makes sense Air Monitoring - Swirl and vortex meters will aid green hydrogen production - Beyond the Stack: Emi...

View all digital editions

Events

Safety & Health Expo 2024

Dec 02 2024 London, UK

Valve World Expo

Dec 03 2024 Dusseldorf, Germany

Aquatech China 2024

Dec 11 2024 Shanghai, China

Abu Dhabi Sustainability Week

Jan 12 2025 Abu Dhabi, UAE

World Future Energy Summit

Jan 14 2025 Abu Dhabi, UAE

View all events