• Precise Water Testing with Minimised Effort

Water Testing

Precise Water Testing with Minimised Effort

The chemical oxygen demand, ST-COD value (ST = small scale sealed tube), of a sample is 
determined by a dichromate method. The result can be considered as an indicator of the theoretical oxygen demand, i.e. the amount of oxygen required for the total chemical oxidation of the 
organic matter present in the water.

The Lovibond COD VARIO Set-Up allows precise water testing with minimised effort. 
The ST-COD concentration in mg/l O2 is measured by photometric detection.

After adding 2 ml of the sample to a Lovibond COD VARIO tube test (LR, MR according to ISO 15705:2002), the tube is heated in the reactor for 2 hours at 150 °C, cooled down and then 
analysed in MD 200 COD VARIO photometer.

The COD Set-Up is a "ready to use" unit with a MD 200 COD VARIO photometer, 25 tube tests for LR and MR (each), the RD 125 thermoreactor for sample digestion and a tube stand.


Digital Edition

AET 28.4 Oct/Nov 2024

November 2024

Gas Detection - Go from lagging to leading: why investment in gas detection makes sense Air Monitoring - Swirl and vortex meters will aid green hydrogen production - Beyond the Stack: Emi...

View all digital editions

Events

Safety & Health Expo 2024

Dec 02 2024 London, UK

Valve World Expo

Dec 03 2024 Dusseldorf, Germany

Aquatech China 2024

Dec 11 2024 Shanghai, China

Abu Dhabi Sustainability Week

Jan 12 2025 Abu Dhabi, UAE

World Future Energy Summit

Jan 14 2025 Abu Dhabi, UAE

View all events