Air Monitoring

Emission Monitoring with Set CEM CERT

Mar 14 2014

Author: Rudi Spinner & Bob Lane on behalf of Siemens

Free to read

This article has been unlocked and is ready to read.

Download

Protection of the environment and human health is a major concern for the European Union. Beside the establishment of different directives (WID, LCPD, IPPC) for the industry, performance criteria and test procedures for automated measuring systems (AMS) have been put in place under EN15267-3. The Siemens Set CEM CERT system has been certified by TUEV Sued in Germany for fulfilling these requirements. Set CEM CERT is used for continuous monitoring of NOx, SO2, CO and O2 in stack gases emitted by power plants, boilers, gas turbines, furnaces, incineration plants, chemical plants, etc. The certificate is the basis for acceptance in all EU member states, and due to the modular concept of Set CEM CERT the system can be applied to various emission monitoring tasks.

Measurement Techniques

The stringent requirements of the European Commission require a very high performance for CEMS (continuous emission monitoring systems). The emission limit values (ELV’s) are becoming lower and hence the performance of the measurement equipment is becoming more important. CEMS can mainly be divided into two categories: extractive or in-situ.  Set CEM CERT (Figure 1) is based on so called “cold-extractive” technology. A heated sample probe with a filter element continuously extracts a sample, after which the heated sample line transfers the sample to a compressor gas cooler, where the moisture content is taken out. The dried gas can then be measured by dedicated analysers. One of the benefits of this cold-extractive technology is that no moisture measurement is required, since the measurement is already based on dry base, as required by the according directives. The components NO, SO2 and CO are determined by a non dispersive infrared (NDIR) analyser (ULTRAMAT 23). For lower emission limit values, gas turbine emission monitoring applications, for example, a non dispersive ultraviolet (NDUV) analyser (SIPROCESS UV600) can be chosen as an alternative. Due to the fact that the emitted gases need to be referenced to a certain oxygen level, an oxygen measurement is required as well, which is offered with either a paramagnetic or electrochemical cell placed in the ULTRAMAT 23.

As mentioned in the preface, there were many test procedures that had to be conducted to the Set CEM CERT during the certification process, in the laboratory as well in the field under real process conditions. Test procedures included checking of parameters such as calibration function, response time, linearity, maintenance interval, zero and span drift, availability, reproducibility, NO converter efficiency, sample flow rate, influence of vibration and, probably most importantly, the influence of specified disturbing and interfering components. Out of the results from lab and field tests the total uncertainty is calculated and listed in the EN 15267-3 certificate.

Before and during the purchase process of Set CEM CERT the operator is required to establish and implement data and information required by EN 14181 in order to assure the quality of the CEMS. The quality assurance process is classified into three different quality assurance levels (QAL’s): QAL 1, QAL 2 and QAL 3. In addition to the QAL-3 step there is an Annual Surveillance Test (AST) to be conducted. AST can be considered as a minor scale of the QAL 2 tests.

Unique System Features

During the field tests of Set CEM CERT according to EN 15267 the influence of each component had to be examined as a whole. For example, it has to be taken into account that SO2 (sulphur dioxide), which is formed when sulphur within the fuel reacts with oxygen, is very water soluble. Due to the cooling of the wet gas in the cooler down to +5°C dew point, the moisture content in the stack gas forms condensate, which then can be removed by peristaltic pumps. If this precaution is not taken, a significant loss of SO2 due to the water solubility in the condensate takes place at low concentration levels. Set CEM CERT offers a solution to prevent this loss of SO2. An acidification module establishes a sufficient acidic environment inside the cooler’s heat exchangers in order to replace the water condensate with weak acid. Due to the acidic environment, the solubility of acidic and water soluble gases is minimised to a negligible extent. A very suitable acid for this purpose is diluted ortho-phosphoric acid (H3PO4), which is injected in front of the primary cooler by a peristaltic pump. All the diluted ortho-phosphoric acid solution mixed with condensate is consecutively removed as acid condensate from the bottom part of the cooler by another peristaltic pump. The acidification module (Figure 2) is recommended when measuring SO2 or NO2 concentrations, which has a similar behavior to SO2, in ranges less than 0-150 mg/m3. Therefore, Set CEM CERT can be used as an EN 15267 approved system for emission monitoring in cases where low SO2 concentrations have to be measured, such as gas turbine applications.

Another important benefit of Set CEM CERT is the high variety of different measuring ranges, which have been tested and approved. In case of SO2 the largest range which can be applied is 0-7000 mg/m3. Furthermore, the modular package allows the certified use of system components from two different leading European manufacturers, thus keeping the number of spare parts in stock to a reasonable amount. Up to two different analysers with different measuring ranges can be configured in the Set CEM CERT, and versions for both indoor and outdoor installation are available. When using electric heaters or cabinet air conditioners the ambient temperature application range can be extended according to the local requirements.

Main applications

Set CEM CERT can be used in a variety of applications such as:

•  Power plants fueled with solid, gaseous or liquid fuels

•  Cement plants and lime kilns

•  Plants for steel and aluminum production

•  Glass melting furnaces

The modular measuring equipment is composed of the sampling probe, the heated sample gas line, a two-stage sample gas cooler, a gas pump and the multi-component analysers, ULTRAMAT 23 and SIPROCESS UV600. The gas path splits in parallel after the sample gas cooler. This supplies each analyser separately with sample gas. One advantage for maintenance is that each analyser can be individually serviced without affecting the other. The sample gas coolers have a moisture alarm in case of malfunction. For additional security, each analyser is also protected by a condensation barrier, which stops the gas flow if a malfunction occurs. This guarantees optimum protection for the heart of the Set CEM CERT - the gas analysers. Semi-automatic calibration with zero and span gases is provided by a 3/2-way solenoid valve installed between the first and second cooling stages. This solenoid valve can also be used for AUTOCAL calibration purposes of the ULTRAMAT 23 (fully automatic timing).

 

 

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

AET 28.2 April/May 2024

May 2024

Business News - Teledyne Marine expands with the acquisition of Valeport - Signal partners with gas analysis experts in Korea Air Monitoring - Continuous Fine Particulate Emission Monitor...

View all digital editions

Events

The World Biogas Expo 2024

Jul 10 2024 Birmingham, UK

ICMGP 2024

Jul 21 2024 Cape Town, South Africa

Australasian Waste & Recycling Expo

Jul 24 2024 Sydney, Australia

Chemical Indonesia

Jul 30 2024 Jakarta, Indonesia

China Energy Summit & Exhibition

Jul 31 2024 Beijing, China

View all events